GENERAL | |
Sound pressure level | < 70 dB |
Working temperature | 10 – 35 °C |
Storage temperature | -25 / 60 °C |
Humidity | 20-80% RH |
Installation category | Category P (according to IEC 1010) |
Pollution degree | Degree 2 (according to IEC 1010) |
Sample capacity | 96 vials of 1.5 ml (STD), 12 vials per segment 72 vials of 4.0 ml (LSV), 9 vials per segment 48 vials of 10ml (Super LSV), 4 vials per segment 160 micro vials of 0.5 ml, 20 vials per segment |
Vial dimensions | Maximum vial height: 47 mm inclusive cap. Minimum vial height: 32 mm inclusive cap. |
Loop volume | 5 – 1000 µl |
Injection volume | 1µl – 1 ml, with 1 µl increment |
Injections per vial | max. 9 (Volumes are programmable for each injection) |
Injection methods | Flushed loop injections Partial loopfill injections µl pick-up injections (without sample loss) |
Needle wash | Programmable: Wash between injections Wash between vials Wash at end of series |
Wash solvent | 250 ml internal wash solvent bottle |
Dispenser syringe | 100 µl, 250 µl (standard), 500 µl ,1000 µl or 10ml (only with userprog) |
Vial detection | Missing vial detection by vial sensor |
Sample access | < 10 seconds |
Headspace pressure | Built-in compressor |
Pre-puncturing septa | With air needle, dual needle action |
Switching time injection valve | electrically < 100 msec |
Piercing precision needle | ± 0.6 mm |
Priority sample | Free programmable |
ANALYTICAL PERFORMANCE | |
Reproducibility | Capped and sealed vials |
•Full loop injections | RSD£ 0.3% |
•Partial loopfill injections | RSD£ 0.5% injection volumes > 5 µl, with headspace pressure on the vial and 30µl pre-flush with air-segment |
•µl pick-up injections | RSD£ 1.0% injection volumes > 5 µl, with headspace pressure on the vial |
ELECTRICAL | |
Power requirements | 115 VAC; ±10%; 50/60 Hz; 250 VA 230 VAC; ±10%; 50/60 Hz; 250 VA |
Fuses | For 115 VAC; two 5.0 AT-fuses (slow, ¼” x 1¼”, UL/CSA) For 230 VAC; two 2.5 AT-fuses (slow, 5 x 20 mm, IEC127) |
COMMUNICATION | |
Outputs | Inject marker (Relay & TTL) Vial marker (Relay & TTL) Labelled vial marker(Relay & TTL) Stop I/O (Relay & TTL) BCD output (TTL) 4 Auxiliary outputs (Relay) 2 Programmable outputs (Relay) Alarm output (Relay) 4 bit time base (TTL) |
Inputs | Next injection input (TTL) Next vial input (TTL) Freeze input (TTL) Stop I/O (TTL) 4 programmable inputs (TTL) |
Serial communication port | RS 232C or Multilink |