Spark Holland Model 900 Triathlon Autosampler

400.00$

1 in stock

SKU: MB0399

Description

Actual item displayed. Comes with 30 day money back guarantee. Specs listed below were taken from manufacturers literature.


GENERAL
Sound pressure level < 70 dB
Working temperature 10 – 35 °C
Storage temperature -25 / 60 °C
Humidity 20-80% RH
Installation category Category P (according to IEC 1010)
Pollution degree Degree 2 (according to IEC 1010)
Sample capacity 96 vials of 1.5 ml (STD), 12 vials per segment
72 vials of 4.0 ml (LSV), 9 vials per segment
48 vials of 10ml (Super LSV), 4 vials per segment
160 micro vials of 0.5 ml, 20 vials per segment
Vial dimensions Maximum vial height: 47 mm inclusive cap.
Minimum vial height: 32 mm inclusive cap.
Loop volume 5 – 1000 µl
Injection volume 1µl – 1 ml, with 1 µl increment
Injections per vial max. 9 (Volumes are programmable for each
injection)
Injection methods Flushed loop injections
Partial loopfill injections
µl pick-up injections (without sample loss)
Needle wash Programmable:
Wash between injections
Wash between vials
Wash at end of series
Wash solvent 250 ml internal wash solvent bottle
Dispenser syringe 100 µl, 250 µl (standard), 500 µl ,1000 µl
or 10ml (only with userprog)
Vial detection Missing vial detection by vial sensor
Sample access < 10 seconds
Headspace pressure Built-in compressor
Pre-puncturing septa With air needle, dual needle action
Switching time injection valve electrically < 100 msec
Piercing precision needle ± 0.6 mm
Priority sample Free programmable
ANALYTICAL PERFORMANCE
Reproducibility Capped and sealed vials
•Full loop injections RSD£ 0.3%
•Partial loopfill injections RSD£ 0.5% injection volumes > 5 µl, with
headspace pressure on the vial and 30µl
pre-flush with air-segment
•µl pick-up injections RSD£ 1.0% injection volumes > 5 µl, with
headspace pressure on the vial
ELECTRICAL
Power requirements 115 VAC; ±10%; 50/60 Hz; 250 VA
230 VAC; ±10%; 50/60 Hz; 250 VA
Fuses For 115 VAC; two 5.0 AT-fuses
(slow, ¼” x 1¼”, UL/CSA)
For 230 VAC; two 2.5 AT-fuses
(slow, 5 x 20 mm, IEC127)
COMMUNICATION
Outputs Inject marker (Relay & TTL)
Vial marker (Relay & TTL)
Labelled vial marker(Relay & TTL)
Stop I/O (Relay & TTL)
BCD output (TTL)
4 Auxiliary outputs (Relay)
2 Programmable outputs (Relay)
Alarm output (Relay)
4 bit time base (TTL)
Inputs Next injection input (TTL)
Next vial input (TTL)
Freeze input (TTL)
Stop I/O (TTL)
4 programmable inputs (TTL)
Serial communication port RS 232C or Multilink

 

Additional information

Weight 60 lbs
Dimensions 30 × 20 × 24 in
Brand

Spark Holland

Model

900

Serial Number